Calibration Products
Made for efficient control and adjustment of the Nilochecker probes

Even the finest equipment needs to be controlled and sometimes adjusted. All our probes have a digital processor that stores calibration data internally. This is why you can continue to work with the NiloChecker, even if a probe is sent for calibration.
Same simple procedure for all probes
One simple adaptor for all probes
Though Nilotech offers calibration service, all our sensors can be calibrated by your local distributor or calibration provider. A universal adaptor is available to control or adjust any NiloChecker probe. More details below.

Gas calibration made easy
With the calibration cap it is easy to calibrate the DG112- CO2 /O2 probe for the NiloChecker. The Calibration cap is a transparent Silicone cap fitting around the top of the DG 112. By using the cap, calibration can be done quickly and precise.

Software for calibration
The intuitive calibration software guides technicians through a control or calibration procedure of any Nilotech measurement probe.

Why Calibration
What is calibration and why is it needed
All electronic components and sensors change properties over time. This is called drift. Though Nilotech only uses the highest quality components and sensors, the drift may result in less precise measurements over time.
Fortunately, all probes have been manufactured so they can be compensated for the drift. During calibration, the output from the probe is compared to a precise reference. Compensation values may be programmed into probe to compensate for the drift.
How often must probes be calibrated
The calibration interval, which is the time between each calibration, depends on the type of probe. Some are more stable over time than others. Typical calibration interval is 12-24 months.
The calibration date and a calibration interval has been programmed into the probe memory by Nilotech.

The calibration house can change this interval and may propose do so, if they see large drift.
Nilotech offers traditional calibration where you get a reminder about when to submit your probe to Nilotech and we return your probe after calibration.
An adaptor will be needed to control or adjust any NiloChecker probe. The Adaptor connects the probe to a Windows PC. Our adaptor can be used with any NiloChecker probe and the intuitive calibration software.
Download Calibration Instructions
Order no: 115s001
With the calibration cap it is easy to calibrate the DG112- CO2 /O2 probe for the NiloChecker. The Calibration cap is a transparent Silicone cap fitting around the top of the DG 112. By using the cap, calibration can be done quickly and precise.
Order no: 115s002
The most important aspect of calibration is to have access to precise references. Nilotech offers reference gasses from Calgaz in 0.9 l cylinders. We supply a 99,999 %N2 gas for 0-calibration and a 5%O2/6%CO2 for span calibration.
Each bottle contains 34l compressed gas for approximately 40 calibrations
Order no: 115s003 (includes one of each bottle)
A precise regulator for the Calibration gas bottles. This regulator gives a constant flow at 0,5 l/min. which ensures correct flow through the calibration cap. (part no. 115s002)
Order no: 115s004
Replacement Oxygen (O2)sensor for Nilotech:
DG112 CO2/O2 probe
DG122 Flow-through CO2/O2 probe.
Estimated lifetime: 2years.
Order no: 112sp001.
Replacement Oxygen (O2)sensor for Nilotech:
DG126 CO2/O2 probe with pump
Estimated lifetime: 2years.
Order no: 126sp001.



Download the instruction on how to use the adaptor and probe calibration software.
Download the instruction on how to use the adaptor and probe calibration software.
NiloChecker
The New all-in-one instrument. Guides you through control and calibration of critical equipment in your laboratory.